EXACTECH, INC.
61 recall(s) on record — most severe on record: Class II.
Recalls by year
- Recalls (61)
Registered facilities (2)
- EXACTECH, INC. FEI 1038671 · FL
- Exactech Sarasota FEI 3008658204 · FL
Recalls & enforcement
20261 record
- Class IIOngoing2026-02-03Impactor handle may be missing cross-pin
20252 records
- Class IIOngoing2025-09-05Due to products that have not completed design updates and performance testing being released/distributed. Products may not conform to design and performance criteria.
- Class IIOngoing2025-08-15Reverse Shoulder humeral liners have an articular surface position outside of the dimensional specification within the labeling.
20244 records
- Class IIOngoing2024-12-31Due to the lack of the Ethylene Vinyl Alcohol (EVOH) barrier layer.
- Class IIOngoing2024-04-26The AcuMatch L-Series 22mm Inner Diameter Bipolar Hip Liner lots were packaged without the specified ethylene vinyl alcohol (EVOH) layer. Between 2004 and August 2021, our packaging process utilized two different types of packaging materials: 1) Low Density Polyethylene (LDPE), N…
- Class IIOngoing2024-04-18Exactech is recalling all affected UHMWPE (ultra-high molecular weight polyethylene) knee patella components packaged in out of specification vacuum bags.
- Class IIOngoing2024-03-06The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH).
20232 records
- Class IIOngoing2023-11-21Exactech has received 3 complaints for 4 devices related to a loss of vacuum in the inner-most vacuum bag.
- Class IIOngoing2023-06-05The ball and spring housed within the locking mechanism of the handle can dislodge during disassembly or during impaction.
20221 record
- Class IIOngoing2022-08-11Specific GXL acetabular polyethylene liners, packaged in non-conforming bags, may adversely impact the device and contribute to accelerated wear.
20214 records
- Class IIOngoing2021-08-30Inserts were packaged in vacuum bags that lacked an additional oxygen barrier layer.
- Class IIOngoing2021-06-29Risk of edge-loading and premature prosthesis wear is possible in a specific subset of patients with certain implant configurations and surgical implant positioning.
- Class IICompleted2021-06-29Biolox Delta Femoral Head was packaged and labeled as 170-36-00 (36mm +0), but the device was marked as 170-36-03 (36mm +3.5).
- Class IICompleted2021-05-11The HAT trial contains a retaining ring that has shown the potential to disengage during use.
20206 records
- Class IIOngoing2020-11-09Potential for incorrect Instructions For Use (IFU) included with an Equinoxe Humeral Stem, Primary, Press Fit, 11mm package.
- Class IITerminated2020-10-06The product may be mislabeled.
- Class IITerminated2020-09-25The device may be incorrectly laser marked as a right stem, when it should have been laser marked as a left stem.
- Class IITerminated2020-08-24Exactech is recalling the Equinoxe Reverse Shoulder Compression Screw/Locking Cap Kit, because it may contain a incorrect length screw.
- Class IIOngoing2020-08-10The Vantage Total Ankle Tibial Insert, Fixed Bearing, Size 4, Right 10mm, were found to have a labeling error. Exactech is recalling the Vantage Total Ankle Tibial Insert, Fixed Bearing, Size 4, Right 10mm because they were found to have a labeling error. The upper right corner…
- Class IIOngoing2020-01-27These Truliant Splined Stem Extension, 18mm x 120mm were found to have a labeling error. The upper right corner of the label incorrectly identifies the implant size as 17mm x 80mm when the implant size is 18mm x 120mm.
20196 records
- Class IITerminated2019-12-03Labels with incorrect translations were identified by Exactech is a distributor in the Netherlands. The translations on the outer labels include the word " long" in all languages other than the original English version. The implants are standard length fracture stems.
- Class IITerminated2019-10-02Potential for the 2.4mm x 2.5 Fluted Pin Pouch to contain a single pin rather than four (4) as intended.
- Class IITerminated2019-09-10Potential for the wire to fracture at the threads and remain in the glenoid.
- Class IITerminated2019-07-24The threaded mechanism of the Threaded Offset Cup Impactor may be gritty and difficult to turn potentially due to entrapped blast media.
- Class IITerminated2019-04-26The Opteform Disc-30mm, Opteform Disc-45mm, and Opteform Disc-90mm were potentially exposed to higher temperatures than those documented in the IFU due to a failure of the cold storage equipment.
- Class IITerminated2019-03-132.5 Truliant Fit Trays were mismarked as 3
20184 records
- Class IITerminated2018-12-05Possibility that the Equinoxe Preserve Humeral Stem, 8mm, was inadvertently packaged with an Optetrak Knee System instructions for use (IFU) rather than the Equinoxe Preserve Humeral Stem IFU as intended .
- Class IIOngoing2018-08-21The Equinoxe Reverse Drill Bit Kits may contain two 2mm diameter drill bits instead of one 2mm diameter drill bit and one 3.2mm diameter drill bit as intended.
- Class IITerminated2018-07-13Exactech is recalling the Tibial Alignment Guide because the Global Trade Item Number (GTIN) on the device is incorrect. The GTIN on the device reads 10885862525629, when it should read 10885862525628.
- Class IIOngoing2018-05-08Potentially have non-conforming internal threads.
201710 records
- Class IITerminated2017-10-09Wrong device description of the label identifying the implant
- Class IITerminated2017-09-19There is a design incompatibility between certain sizes of the offset spine stiffener screw and the constrained condylar tibial insert when used in conjunction with an offset tibial tray.
- Class IITerminated2017-08-31Potentially mislabeled.
- Class IITerminated2017-07-11Vix Bit may fracture during use.
- Class IITerminated2017-06-23The Truliant Tibial Trial Handle's pin may disassociate from the main body.
- Class IITerminated2017-05-25Instructions for use (IFU) is updated to clarify the proper technique and instrumentation to ream the Glenoid, i.e., to include a caution statement that the surgeon should ream on-axis and that reaming off-axis may result in instrument breakage.
- Class IITerminated2017-05-25The firm determined that the labeling (surgical technique) should be updated to clarify the proper technique and instrumentation to remove the tamp assembly from the bone, i.e., to include a caution statement about the potential for instrument breakage if the tamp handle/guide is…
- Class IITerminated2017-05-22The firm determined that the labeling (surgical technique) should be updated to clarify the proper technique and instrumentation to perform calcar planing to the bone, i.e., to include a caution statement about the potential for instrument breakage if bending or torsional forces…
- Class IITerminated2017-05-16Possible inner labeling and packaged device not matching the outer label on the box.
- Class IITerminated2017-04-07Missing Screw in carton. Nine (9) serial numbers were identified as being inadvertently packaged as one per carton when two per carton are required.
20164 records
- Class IITerminated2016-11-07Products were labeled with incorrect Global Trade Item Numbers. The GTIN on the label identifies the devices as another.
- Class IITerminated2016-10-25Disengaged screw head could potentially not be retained in the Torque Defining Screwdriver.
- Class IITerminated2016-05-26Labeled with the incorrect Global Trade Item Number (GTIN). The GTIN on the label identifies the devices as another.
- Class IITerminated2016-04-12Devices were inadvertently sent to packaging and finished goods prior to the PMA site-change supplement approval.
20154 records
- Class IITerminated2015-09-28Incorrect external and internal packaging labels. Exactech has identified nineteen (19) Acapella One Interbody Fusion Device Lordotic (05-090-01-1610) were packaged as Acapella One Interbody Fusion Device Parallel (05-090-02-1610)
- Class IITerminated2015-09-23Eight (8) pieces did not receive the specified Hydroxyapatite (HA) coating.
- Class IITerminated2015-05-21Shipped to a single consignee prior to completion of final inspection.
- Class IITerminated2015-02-13Incorrectly packaged. Outer and inner labeling may incorrectly identify the size of the enclosed device.
20143 records
- Class IITerminated2014-10-09Mislabeled
- Class IITerminated2014-03-07The femoral stem is lacking specified plasma coating.
- Class IITerminated2014-01-06The products have been identified to contain springs which have been manufactured with type-304 stainless steel rather than the specified titanium alloy.
20133 records
- Class IITerminated2013-09-11Exactech is recalling the InteGrip Acetabular Augments due to an out of range condition for an in-vitro biological evaluation standard.
- Class IITerminated2013-05-30When torque is applied to the Articulated Driver at the maximum angle, the cross pin may shift out of position, potentially leading to disassembly of the instrument.
- Class IITerminated2013-02-01A dimensional mismatch was identified which can potentially allow for interference between devices.
20123 records
- Class IITerminated2012-11-05Recall of Novation Crown Cup Connection GXL Neutral Liners Catalog No. (130-32-53) and Novation Crown Cup Connection GXL + 5mm Lateralized Liners Catalog No. (136-32-53). Exactech discovered the device labeling was inadvertently transposed.
- Class IITerminated2012-06-15Exactech, Inc. initiated a recall of Equinoxe Replicator Plates, Equinoxe Locking Screws, Equinoxe Compression Screw Kits, Equinoxe Fracture Adapter Tray Screws, and Octane-C Intervertebral fusion devices due to the firm determining that the device labeling identifies the affecte…
- Class IITerminated2012-01-24Exactech Inc. of Gainesville, FL is recalling their Novation Modular Drill Bits after the devices were reported to have been used by end user while in a non-sterile condition.
20113 records
- Class IITerminated2011-10-17Exactech, Inc. of Gainesville, FL recalled their Optetrak Proximal Tibial Spacer after the devices were determined to be dimensionally incompatible to mate as intended with the Logic Tibial Tray.
- Class IITerminated2011-09-23Exactech, Inc. Gainesville, FL is recalling their Optetrak Cemented Trapezoid Tibial Tray, Size 1F/1T, 2F/1T and the Optetrak Rotating Bearing Knee Tibial Tray, Finned, Cemented Tibial Tray, Size 2F/1T. Both trays have the potential to be mislabeled.
- Class IITerminated2011-08-19Exactech, Inc. of Gainesville, FL is recalling their Optetrak Logic Posterior Stabilized Tibial Insert, Size 6 after the device was determined to be out of dimensional specification.
20091 record
- Class IITerminated2009-03-06Incorrect thread form on the screw.
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. We restate the public record only — measured facts, never risk verdicts.
Cite this record
RegWarden. “EXACTECH, INC. — FDA compliance profile.” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/company/oo-1038671-exactech-inc/
Alerts for EXACTECH, INC.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.