Noah Medical
5 recall(s) on record — most severe on record: Class II.
Recalls by year
- Recalls (5)
Registered facilities (4)
- NOAH MEDICAL Corp. FEI 3033628477 · CA
- NOAH MEDICAL Corp. FEI 3021325287 · CA
- NOAH MEDICAL Corp. FEI 3027205723 · CA
- Tianjin Puli Ark Medical Technology Co., Ltd FEI 3025749544 · CN
Recalls & enforcement
- Class IIOngoing2025-09-12Due to Users not having access to the Instructions for Use/User Manual due to it being password protected.
- Class IIOngoing2024-10-28Bronchoscopic visualization system power supplies have defective semiconductor components that may fail unexpectedly, which could result in an aborted procedure since it is a non recoverable failure. Manual removal of the bronchoscope when the system is not functioning can increa…
- Class IIOngoing2024-10-25Bronchoscopic system uninterruptable power may have defective component, may cause power supply short circuit/trip protective circuit breakers causing sudden, unrecoverable power loss, which could occur while system is on mains or internal battery power; if during procedure, coul…
- Class IIOngoing2024-10-25Not all biopsy tools meant to fit bronchoscope channel (2.1mm inner diameter/length greater or equal to 1100mm) will pass through if scope channel positioned in sharp bend (approx. greater than 100-deg.), excessive force may result in tissue injury/pneumothorax; instruments with …
- Class IIOngoing2024-03-09Due to a misaligned calibration of the wired controller that may result in unintended motion of the scope.
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. We restate the public record only — measured facts, never risk verdicts.
Cite this record
RegWarden. “Noah Medical — FDA compliance profile.” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/company/oo-10087546-noah-medical/
Alerts for Noah Medical
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.