Kelyniam Global Inc.
1 recall(s) on record — most severe on record: Class II.
Recalls by year
- Recalls (1)
Registered facilities (1)
- KELYNIAM GLOBAL INC. FEI 3009424796 · CT
Recalls & enforcement
- Class IITerminated2017-06-10Significant design changes made to the Custom Skull Implant (CSI) devices, such as removing pressure relief holes across the length of the device and adding suture holes along the ridge of the implant, were not validated or submitted to FDA for marketing clearance. The Temporal S…
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. We restate the public record only — measured facts, never risk verdicts.