AB SCIEX
6 recall(s) on record — most severe on record: Class II.
Recalls by year
- Recalls (6)
Registered facilities (1)
- AB SCIEX LP FEI 3008336863 · CA
Recalls & enforcement
- Class IIOngoing2023-08-30In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer …
- Class IITerminated2021-10-04The values of the Internal Standard (IS) concentrations are incorrectly derived when the user builds a customized test, then set the IS column to hidden and left the column empty. This may lead to the user making incorrect conclusions using incorrect results.
- Class IITerminated2016-06-16MultiQuant MD software where under certain conditions a user can be presented with incorrect quantitative results when using the Sum Multiple Ions feature.
- Class IITerminated2016-03-25Under certain conditions, mass spectrometers may report incorrect quantitative results.
- Class IITerminated2016-02-24Wrong quantitative results may be displayed in a report from the device, which may potentially lead to an incorrect patient diagnosis.
- Class IITerminated2014-07-15Software- AB SCIEX API 3200MD LC/MS/MS SYSTEM with MultiQuantMD 3.0 software may provide incorrect data for patient diagnosis
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. We restate the public record only — measured facts, never risk verdicts.
Cite this record
RegWarden. “AB SCIEX — FDA compliance profile.” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/company/oo-10041672-ab-sciex/
Alerts for AB SCIEX
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.