ImaCor Inc
2 recall(s) on record — most severe on record: Class II.
Recalls by year
- Recalls (2)
Registered facilities (1)
- IMACOR, Inc FEI 3006382185 · NY
Recalls & enforcement
- Class IIOngoing2025-08-29Production assembly error resulting in the potential for reversed articulation, where moving the lever forward caused the probe tip to anteflex rather than retroflex, and vice versa.
- Class IIOngoing2024-09-16Epoxy used to seal the edge of the transducer into its paddle exceeded specification. This variability could attenuate the ultrasound signal with a potential impact on the quality of images obtained from the probes.
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. We restate the public record only — measured facts, never risk verdicts.